Mts Model FlexTest SE Manual do Utilizador Página 93

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 264
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 92
Model 493.21B (DUC B) Jumpers
Models FlexTest® IIm/GT/SE Controller Hardware FlexTest GT Controller Connections
93
4- or 8-wire
sensor cable
Jumpers W2 and W3 establish the correct shunt calibration connections for 4-
wire and 8-wire sensor cables. The length of the sensor cable determines which
type of cable is used. Short cables are typically 4-wire cables while long cables
are typically 8-wire cables.
Jumper W2 selects a 4- or 8-wire transducer for positive feedback.
Jumper W3 selects a 4- or 8-wire transducer for negative feedback.
The setting of jumper W8 (single ended excitation enable) may affect your
choices for these jumpers (W2 and W3).
Note +/– excitation (–excitation is standard).
Bridge balance Jumper W4 changes the offset of the bridge balance circuit. If the zero offset is
too large when the sensor bridge is balanced, you should enable this feature (it
adds a 100 kΩ resistor to the input of the conditioner preamp).
W2
W3
W2
W3
W2
W3
W2
W3
123
–EX to –FB shunt
in 4-wire config
-EX to +FB shunt in
4-wire config
–EX to –FBR shunt
in 8-wire config
-EX to +FBR shunt
in 8-wire config
Default
Default
W4
W4
Disable
Enabl
Disable
Vista de página 92
1 2 ... 88 89 90 91 92 93 94 95 96 97 98 ... 263 264

Comentários a estes Manuais

Sem comentários