A running max/min meter provides the history of sensor extremes that the meter has read since it was
last reset.
4. Set up a Running Max/Min Force Meter.
Use the following parameters:
• Meter Type — Running Max/Min
• Channel — Ch1 (or other depending on your configuration)
• Signal — Force
• Engineering Units — kN
5. Set up a Timed Displacement Meter.
Use the following parameters:
• Meter Type — Timed
• Channel — Ch1 (or other depending on your configuration)
• Signal — Displacement
• Engineering Units — mm
A timed meter provides the current reading of the sensor.
Important: Align the meters as shown in the preceding illustration by clicking the Align button.
You can cycle through the available alignments by clicking multiple times.
6. Set up a Timed Force Meter.
MTS TestLine Test System Operation | 79
Running the Example HCF Test