Parameter sets are found in the lower left corner of the Open Station window. For this example HCF
test, select the HardSpecimenTuning parameter set and click Open. A set of windows may open
based on the last user’s window setup. Minimize any windows that come up, as you will be bringing
them up as appropriate.
6. Ensure that the HSM (hydraulic service manifold) is off.
Turning the HSM off removes power from the actuator so that it cannot move unexpectedly.
Note: It is acceptable to leave the pump (HPU) on as it may be needed for other stations.
7. Identify the current configuration and parameter set.
The configuration and parameter set currently in use is shown in the title bar of the Station Manager
application. For this test, the title bar should read "Training config.cfg: hard specimen tuning".
Verify Calibration Files
1. Verify that the correct calibration file for the displacement sensor is in use.
MTS TestLine Test System Operation | 53
Running the Example HCF Test