Model 494.42 Single-Station System Board
MTS FlexTest® Models 40/60/100/200 Controller Hardware FlexTest Controller Configurations
67
Model 494.42 Single-Station System Board
About the Model 494.42 Single-Station System Board
The Model 494.42 System Board is a two-
slot board that is only used in the Model
494.04 Chassis. This board provides analog
and digital I/O, E-Stop, HSM, UPS, and
HPU control for one station.
Board features
External Interfaces
Three digital inputs (J54)
Three digital outputs (J55)
Auxiliary power outputs for external
devices (J49)
UPS inputs (J56) ACFail and low
battery inputs (switch contacts)
Two analog outputs (DA Output
connector on the 494.04 chassis)
Hydraulic Control
24-volt HPU control (J25)
24-volt HSM control (Off/Low/High or
proportional) for one HSM (J28)
E-Stop/Interlock Control
Load-frame E-Stop and crosshead
control for one load frame (J29)
Optional E-Stop inputs (J24-located on
the 494.04 chassis)
Interlock (J43) provides one interlock
input, one interlock output, and a
software-defined event input
E-Stop/Run outputs (J23)
Power
100-240 VAC
50-60 Hz, 1-2 A
DA Output
J24
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8
LAN 2 LAN 1
DEBUG
MOTOROLA
10/100 BASE T 10/100 BASE T
SCSI
BUSY
PIB
BUSY
PCI MEZZANINE CARD
PCI MEZZANINE CARD
J54 Dig InJ55 Dig Out
J43 INTLKJ56 UPS
J49 Aux Pwr J23 Estop/Run
J29 Load FrameJ25 HPU
J28 HSM
J54 Dig In J55 Dig Out
J43 INTLK J56 UPS
J49 Aux Pwr J23 Estop/Run
J29 Load Frame J25 HPU
J28 HSM
J49 AUX PWR
Auxiliary DC supply
voltages for external
devices:
+ 5 V DC, +15 V DC,
-15 V DC, +24 V DC
J28 HSM
24-volt control for
HSM off/low/high or
proportional solenoid
plus an HSM Estop
input
J43 Intlk
Interlock input
Interlock output
Event input
J29 Load Frame
Load Frame E-Stop
J25 HPU
24 V DC HPU control
for one HPU channel
J54 Dig In
Digital inputs (3)
J55 Dig Out
Digital outputs (3)
Model
494.04
Chassis
J23 Estop/Run
2 Estop NO outputs
1 set of Run NO/NC
outputs
DA Output (2))
J24 E-Stop
(Optional)
J56 UPS
UPS digital inputs
ACFail, Low battery
Vista de página 66