Model 494.21 Multi-Range DUC with Acceleration
MTS FlexTest® Models 40/60/100/200 Controller Hardware Mezzanine Cards
173
Model 494.21 Multi-Range DUC/Acceleration Compensation Card Settings
Software configuration Most configuration and setup of this card is done through software.
Note For detailed information on software configuration, calibration, and
tuning, see the control software and tuning/calibration manuals.
Hardware settings The conditioner portion of this card has a two-position switch (SW1) that is
used to set up the following DC conditioner functions:
• Switch 1A connects a bridge-completion resistor to -FBR to complete a
quarter-bridge circuit.
• Switch 1B provides a ground reference for an external transducer by
connecting -FBR to ground.
Note Most hardware settings are factory installed. For custom
applications, refer to the system configuration drawings for analog
bus routing, interconnects, cable part numbers, and other
information.
Switch Settings
A = Bridge Completion Switch
The ON position connects a
user-installed bridge completion
resistor to complete a quarter-
bridge circuit.
B = Grounding Switch
The ON position connects -FBR
line to ground.
SW1