Waste Electrical and Electronic Equipment (WEEE) Considerations
The Waste Electrical and Electronic Equipment (WEEE) symbol ( )
indicates that the controller and its electronic parts must not be disposed of as
unsorted municipal waste. Proper disposal is required by approved electronic
waste collection agencies. Customers in the EC region who desire to return an
end-of-life controller and its electronic parts are encouraged to contact your local
MTS Systems Sales/Service Offices for instructions.
Electrical surge immunity test EN 61000-4-5 2 kV mains line to earth
1 kV mains line to line
.5 / 1 kV cables >30 meters line to earth
Performance Criteria B
Radio frequency common mode,
amplitude modulated
EN 61000-4-6 10 V (rms)
Performance Criteria A
Power frequency magnetic field EN 61000-4-8 30 A (rms)/m 50 and 60 Hz
Performance Criteria B
Voltage dips, short interruptions,
and voltage variations
EN 61000-4-11 1 cycle and 0% amplitude of cycle
Performance Criteria B
10 cycle and 40% amplitude of cycle
25 cycle and 70% amplitude of cycle
250 cycle and 0% amplitude of cycle
Performance Criteria C
Radiated emissions EN 55011 Class A
Conducted emissions EN 55011 Class A