Concentric Adjustments
709 Alignment System Alignment Wizard
111
If the upper gauge X coordinate is positive
1. Initially loosen C270.
2. Tighten C90 moving the red (upper) and blue (lower) gauge readings
together until they slightly pass each other.
3. Tighten C270 to bring the red and blue gauge readings back to
vertically align with each other as close as practical.