Mts Material Test Systems Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Hardware Mts Material Test Systems. MTS Material Test Systems User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - 100-196-370 D be certain

MTS Material Test System Operation100-196-370 D be certain.

Página 2

2.1.0 How to Get Technical SupportStart with your manualsThe manuals supplied by MTS provide most of the information you need to use and maintain your

Página 3 - Contents

Know information from prior technical assistanceIf you have contacted MTS about this problem before, we can recall your file based on the:• MTS notifi

Página 4 - 9.0 Decommission 75

2.3.0 If You Contact MTS by PhoneA Call Center agent registers your call before connecting you with a technical support specialist. The agentasks you

Página 5 - 1.0 Preface

After you callMTS logs and tracks all calls to ensure that you receive assistance for your problem or request. If you havequestions about the status o

Página 7

3.0 IntroductionTopics:•Manual Overview...

Página 8

3.1.0 Manual OverviewThis manual provides operating guidelines, installation guidelines, and safety practices for operation of anMTS Test System. This

Página 9 - 2.0 Technical Support

DescriptionItemTest Controller Software (Typical)1• 793 System Software• 793 Applications• 793 UtilitiesCables2Transducer ID Cable3Outlet Strip4Line S

Página 10 - 2.2.0 Before You Contact MTS

Figure 3: Hydraulic Service ManifoldTable 2: Hose connections from the HPU and to the actuatorsDescriptionItemPressure1Return2Drain3Figure 4: Load Fra

Página 11

DescriptionItemHydraulic service manifold2Hose connections:3• Pressure• Return• DrainHand set (some models)4Actuator speed switch (some models)5E-Stop

Página 12

© 2013 MTS Systems Corporation. All rights reserved.MTS, TestStar, and TestWare registered trademarks of MTS Systems Corporation within the United Sta

Página 13

DescriptionItemEnvironmental Accessories11A. High temperature furnace1B. Environmental chamber1C. Temperature controllerExtensometers2Alignment access

Página 14

3.2.2.0 Servohydraulic and Mechanical ComponentsThe following paragraphs briefly describe the servohydraulic and mechanical components that are typica

Página 15 - 3.0 Introduction

Hydraulic power supplyThe hydraulic power unit (HPU) provides pressurized hydraulic fluid to the servovalve. An HPU typicallyincludes a reservoir for

Página 16 - 3.1.0 Manual Overview

Table 3: Simple Control diagramDescriptionItemTest command1Summing and conditioning2Station manager3Closed loop4Feedback source5When reduced to its ba

Página 17

3.4.0 InterlocksThe test controller incorporates interlock circuitry that can automatically stop a test when a system problem,specimen failure, or an

Página 18

4.0 SafetyTopics:•Overview...

Página 19

4.1.0 Overview4.1.1.0 Safety Information OverviewMTS systems are designed to generate single-axis or multi-axial motions and forces simultaneously in

Página 20

4.2.0 Safety Practices4.2.1.0 General Safety PracticesIf you have system related responsibilities (that is, if you are an operator, service engineer,

Página 21

Locate lockout/tagout pointsKnow where the lockout/tagout point is for each of the supply energies associated with your system.Thisincludes the hydrau

Página 22 - 3.3.0 System Control Theory

Ensure Correct Cable ConnectionIf a system cable has been disconnected, ensure that you establish the correct cable-to-connector relationshipduring re

Página 23

Contents1.0 Preface 51.1.0 Before You Begin...

Página 24 - 3.4.0 Interlocks

the recommended procedure for bleeding a pneumatic or gas-charged device, because it can expose youto the dangers of escaping compressed gas and parti

Página 25 - 4.0 Safety

Provide test area guardsUse protective guards such as cages, enclosures, and special laboratory layouts when you work with hazardoustest specimens (fo

Página 26 - 4.1.0 Overview

Customer must evaluate risks due to ejected parts or materials from the test specimens. If the MTS Test AreaEnclosure option is not selected by the cu

Página 27 - 4.2.0 Safety Practices

Contain small leaksDo not use your fingers or hands to stop small leaks in hydraulic or pneumatic hoses. Substantial pressurescan build up, especially

Página 29

5.0 Installation GuidelinesTopics:•Installation Guidelines...

Página 30

5.1.0 Installation GuidelinesThe MTS Test System is typically installed by qualified MTS field service engineers. This section providesreference infor

Página 31

laboratory personnel, use of an overhead crane, lifting straps or a forklift might be necessary to handlecomponents or fixtures.Heat dissipationTo pro

Página 32

addition, it is recommended that the hydraulic power unit be located in a room separate from the test system,if possible, to reduce noise in the test

Página 33

must be provided by the customer to allow maintenance personnel to safely remove all power to the HPU.Wiring must conform to local electrical codes an

Página 34

5.7.0 Connecting the System Components...405.8.0 System Star

Página 35 - 5.0 Installation Guidelines

Due to the high center of gravity of the console, two people are required to roll the console if the floor hasobstructions or bumps; use one person on

Página 36 - 5.2.0 Facility Preparation

supplied with the system, some of these connections might not apply. If provided, refer to the system assemblyand/or hydraulic distribution drawings (

Página 37

ActuatorsAfter checking the electrical and hydraulic power unit connections, the actuator position control electronics(the Set Point control on the se

Página 38

6.0 Getting StartedTopics:•Overview...

Página 39 - 5.4.0 Grounding Requirements

6.1.0 OverviewBefore setting up or operating the test system, the hardware component configuration must be defined, controlsettings (or test parameter

Página 40

For balanced system operation, the load frame crosshead should be positioned so that the actuator pistonrod will be at approximately midstroke when th

Página 41 - 5.8.0 System Startup

than one signal is selected as the input, the underpeak detector will not trigger an interlock unless all of theselected signals are below the specifi

Página 42

Table 5: Effects of Span and Setpoint on Program SignalDescriptionItemFull-scale operating range1Testing range2Span affects program amplitude3Setpoint

Página 43 - 6.0 Getting Started

Table 6: Error detection levelDescriptionItemProgram command1Feedback2DC Error3Time lag4The error detector can be preset to detect an excessive DC err

Página 44 - 6.1.0 Overview

through the system interlock. Likewise, if the feedback falls below the preset lower limit detect level, the limitdetector will (if enabled during tes

Página 45 - 6.2.3.0 Set Up the Counter

1.0 PrefaceTopics:•Before You Begin...

Página 46 - 6.3.2.0 Define Test Program

Determining limits from known test forces or displacementsIf the maximum and minimum test forces or displacements are known, the limit detector levels

Página 47

Table 8: Underpeak Detector LevelsDescriptionItemFull-scale operating range1Maximum underpeak detect level2Maximum underpeak detected3Minimum underpea

Página 48

• The input to the underpeak detector is the transducer feedback signal from a load cell with a full-scaleoperating range of ±25 kN• There is zero loa

Página 49

Table 9: Gain adjustmentDescriptionItemResponse time1Program2Transducer feedback3Increasing the gain decreases the stability margin of the system, inc

Página 50

DescriptionItemProgram3Transducer feedback4If the gain is set too high, unstable system operation can result.This instability may cause specimen damag

Página 51

SummaryServo loop adjustments are experimental and the operator should become familiar with the effects of theseadjustments while using dummy specimen

Página 53

7.0 OperationTopics:•Operation Overview...

Página 54

7.1.0 Operation OverviewThis section contains a typical operating sequence that will lead you through local test setup and operation.Detailed informat

Página 55

• Determine the position of the load frame crosshead.• Calculate the upper and lower limit detection levels to be used during specimen installation.•

Página 56

1.1.0 Before You BeginSafety first!Before you use your MTS product or system, read and understand the safety information provided with yoursystem. Imp

Página 57 - 7.0 Operation

7.4.0 Adjust Displacement Servo Loop (optional)Note:If the test is to be run in displacement control, then displacement control loop should be tuned f

Página 58 - 7.2.0 Initial Setup

Warning:After you turn off system hydraulics, do not come near the load frame for at least 30 minutes– actuator motion is still possible.Unexpected ac

Página 59 - 7.3.0 Set Detector Levels

Caution:Oil spillage can create an environmental concern and a slippery surface.Slippery surfaces can cause you to fall and incur personal injury.When

Página 60

Note:If possible, install a dummy specimen at this time. Use of a dummy specimen will help avoid potentialdamage to the test specimen when establishin

Página 61 - 7.5.0 Install the Specimen

Caution:Oil spillage can create an environmental concern and slippery surface that can cause personalinjury.When changing hydraulic grips, make sure y

Página 62 - Warning:

Note:If possible, install a dummy specimen at this time. Use of a dummy specimen will help avoid potentialdamage to the test specimen when establishin

Página 63

Note:The following steps are provided as a guide to the servo loop adjustments. Adjustment of these controlswill vary depending on the test specimen a

Página 64

7.7.0 Set Up the Program, Detectors, Counter,and Data AcquisitionFollowing is a brief procedure to set up the program command. Detailed information on

Página 65 - 7.6.0 Adjust Load Servo Loop

9. If a dummy specimen was installed (as recommended in the specimen Installation section, remove dummyspecimen and install test specimen.10. If desir

Página 66

8.0 Routine MaintenanceTopics:•Maintenance Overview...

Página 67

Other special text conventionsImportant:Important notices provide information about your system that is essential to its properfunction. While not saf

Página 68 - 7.8.0 Run the Test

8.1.0 Maintenance OverviewCaution:High pressure fluids can be present in the test system.High-pressure hydraulic fluid can cause cuts and burns.Ensure

Página 69 - 8.0 Routine Maintenance

As a guideline, for a test system that operates eight hours per day, one week will equal 40 hours, and onemonth will equal 160 hours. For a test syste

Página 70 - 8.1.0 Maintenance Overview

5 +Years2 Year1 Year6months3months2months1Month1WeekDailyCalendar time10,0004,0002,0001000500320160408Running time - HoursXInspect for leaks andwearMT

Página 71

5 +Years2 Year1 Year6months3months2months1Month1WeekDailyCalendar time10,0004,0002,0001000500320160408Running time - HoursMTSXCheck functionality ofwa

Página 72

5 +Years2 Year1 Year6months3months2months1Month1WeekDailyCalendar time10,0004,0002,0001000500320160408Running time - HoursXInspect furnace sealsXInspe

Página 73

9.0 DecommissionTopics:•Decommission Test System...

Página 74

9.1.0 Decommission Test SystemThe decommissioning process is performed when the system is going to be moved or taken out of service.Disassembly is req

Página 75 - 9.0 Decommission

Warning:Warning:Residual hydraulic pressure can produce a high pressure spray.This pressurized hydraulic spray can cut you or the hydraulic fluid coul

Página 76

Caution:Warning:Electrical components and parts can contain hazardous chemicals andcompounds.Improper disposal of hazardous chemicals and compounds ca

Página 77

100-196-370 D

Página 80

2.0 Technical SupportTopics:•How to Get Technical Support.......................

Comentários a estes Manuais