Connecting Optional Transformers
Every MTS Criterion Series 40 Material Test System has an optional power
transformer. Customer should select a transformer which supplied by MTS
according to the local voltage range.
The following table provides the transformer requirements for various load frame
types.
Transformer
Connections
The installation procedure for connecting transformers for units is provided
following the figure.
Transformer Selection table
M
ODEL TRANSFORMER PARTNUMBER
TRANSFORMER PARAMETER
C41
C42
100-248-279 1.5 kVA
120 to 220 VAC
C43 (10 kN)
C44 (10 kN)
100-248-281 3.0 kVA
120 to 220 VAC
C43 (30 kN, 50 kN)
C44 (30 kN)
C45 (50 kN)
100-248-281 3.0 kVA
120 to 220 VAC
C45 (100 kN) 100-248-280 5.0 kVA
120 to 220 VAC
C45 (300 kN,600 kN) 100-249-050 14.0 kVA
208 to 480 VAC
C45 (300 kN,600 kN) 100-249-048 14.0 kVA
600 to 480 VAC