Electrical Power Requirements
MTS Criterion
®
Series 40 Site Preparation Guide
System Specifications
21
Electrical Power Requirements
The following table provides the electrical power requirements of various load
frame types.
Note Electrical connections must be made by qualified personnel and conform to local
codes and regulations. Local electrical codes supersede any information found
here.
Electrical Power Requirements
M
ODEL VOLTAGE FREQUENCY CURRENT MAXIMUM POWER
R
EQUIREMENT
C41
C42
Single phase
200-230 VAC
50/60 Hz 5 A 1000 W
C43 (10 kN)
C44 (10 kN)
Single phase
200 - 230 VAC
50/60 Hz 10 A 2000 W
C43 (30 kN, 50 kN)
C44 (30 kN)
C45 (50 kN)
Single phase
200 - 230 VAC
50/60 Hz 12 A 2400 W
C45 (100 kN) Single phase
200 - 230 VAC
50/60 Hz 22 A 4400 W
C45 (300 kN) Three phases
380-480 VAC
50/60 Hz 20 A 5200 W
C45 (600 kN) Three phases
380-480 VAC
50/60 Hz 20 A 5200 W
Facility Power
Requirements:
Provide a dedicated circuit for the machine, computer, and monitor.
Insulation Over Voltage Category II
Pollution Degree 2
Comentários a estes Manuais