Mts Advantage Wedge Action Grips Manual do Utilizador Página 16

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 34
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 15
Manual Name
Introduction
16
Grip Environment
The grips can be used at temperatures from -129 °C to +315 °C
(-200 °F to +600 °F). For testing at temperatures other than ambient conditions,
the grips can be mounted inside an environmental chamber. When using the grips
above ambient temperatures, the maximum tensile load must be reduced
according to the figure below. Also, if the grips are being used above +260 °C
(+500 °F), the stainless steel springs should be removed or they might have to be
replaced periodically.
Vista de página 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 33 34

Comentários a estes Manuais

Sem comentários